BI‑Vital 5.2

Power estimation

Measured currents for 4.2V battery supply, FW 2.0, Default Config, Battery 150 mAh / 0.56 Wh

Device Mode Current Calculated Power* Calculated Runtime max** Calculated Runtime min***
Shipping 8.94 µA 37.5 μW 699 days 621 days
IDLE 8.315 mA 34.9 mW 18 hours 16 hours
Record** 16.58 mA 69.6 mW 9 hours 8 hours

* P=Measured Current * 4.2V
** 150 mAh / Measured Current
*** 0.56 Wh / Power
**** Depends heavily on the recording configuration

Recording time

Estimated recording time with an empty device memory, FW Version ≥ 3.0

Configuration Update rate Recording time
Heart rate
Euler 3-axis
Asynchronous
50 Hz
31h
ECG raw
Acc+Gyr+Mag 3-axis
200 Hz
100 Hz
8h
Heart rate
Acc 3-axis
Temperature, Humidity, Pressure
Asynchronous
100 Hz
1/min
16h