BI‑Vital 5.2
Power estimation
Measured currents for 4.2V battery supply, FW 2.0, Default Config, Battery 150 mAh / 0.56 Wh
| Device Mode | Current | Calculated Power* | Calculated Runtime max** | Calculated Runtime min*** |
|---|---|---|---|---|
| Shipping | 8.94 µA | 37.5 μW | 699 days | 621 days |
| IDLE | 8.315 mA | 34.9 mW | 18 hours | 16 hours |
| Record** | 16.58 mA | 69.6 mW | 9 hours | 8 hours |
* P=Measured Current * 4.2V
** 150 mAh / Measured Current
*** 0.56 Wh / Power
**** Depends heavily on the recording configuration
Recording time
Estimated recording time with an empty device memory, FW Version ≥ 3.0
| Configuration | Update rate | Recording time |
|---|---|---|
| Heart rate Euler 3-axis |
Asynchronous 50 Hz |
31h |
| ECG raw Acc+Gyr+Mag 3-axis |
200 Hz 100 Hz |
8h |
| Heart rate Acc 3-axis Temperature, Humidity, Pressure |
Asynchronous 100 Hz 1/min |
16h |